The CA-2500 2D Color Analyzer is an innovative instrument capable of efficiently measuring the distribution of luminance and chromaticity of multiple display devices simultaneously.

2D Color Analyzer CA-2500

The CA-2500 2D Color Analyzer is an innovative, fourth generation, 2D Color Analyzer from Konica Minolta.  It is also the successor to the CA-2000 2D Color Analyzer. The CA-2500 is an ideal instrument for evaluating luminance and chromaticity mura (non-uniformity) of a variety of display devices, such as LCD and OLED displays, and other light emitting devices such as automobile instrument panels and illuminated keypads. Unlike conventional imaging colorimeters that use RGB color separation filters, the CA-2500 uses XYZ filters that closely match the CIE 1931 color matching functions.  This provides luminance and chromaticity measurements that correlate well with human eye sensitivity.

The CA-2500 2D Color Analyzer now comes with an improved minimum measurable luminance range (0.05 cd/m²) and is more durable, with service life measurement cycles of approximately 5 times that of the predecessor model. Furthermore, the new Data Management Software CA-S25w (included as a standard accessory of CA-2500) provides simple measurements. This is made possible by the introduction of the focus assist function, positioning assist function, and the automatic measurement area extraction function in the software. The software is also capable of measuring multiple samples simultaneously and provides evaluation results for each individual sample, resulting in greater measurement efficiency.

 

2D Color Analyzer CA-2500

  • XYZ filters provide high correlation with the spectral response of the human eye – uses XYZ filters that closely match the CIE 1931 color-matching functions to provide luminance and chromaticity measurements.
  • Expanded low-luminance measurement range – minimum measureable luminance has been Improved from 0.1 cd/m² to 0.05 cd/m².
  • Improved durability – service life measurement cycles have been increased to approximately 5 times that of the CA-2000
  • Interchangeable lenses for measurements of various subjects are available.
  • Advanced Data Management Software CA-S25w, as a standard accessory – focus assist function, positioning assist function, and automatic measurement area extraction function greatly simplify troublesome measurement preparations.
  • SDK (software Development Kit), as a stand accessory, used by customer to efficiently create their own software for controlling the CA-2500.
  • Capable of simultaneous measuring of multiple samples.

2D Color Analyzer CA-2500

ModelCA-2500S
Light ReceptorCCD image sensor (monochrome) ; 2/3-inch ; Effective number of pixels : 1,000 x 1,000 pixels ; Equipped with XYZ filter (closely matches CIE 1931 color-matching function) and ND filter
LensInterchangeable Standard, wide, and telephoto lenses; low-magnification and high-magnification macro rings (for use with telephoto lens)
Measurement Points (Resolution)980 x 980 (490 x 490 or 196 x 196 selectable by using Data Management Software CA-S25w)
Color Indication ModesXYZ, LVxy, LVu’v’, T∆uv, Dominant wavelength, Excitation purity, LV contrast
Display ModesPseudocolor, Chromaticity diagram, Spot, Cross section, Color Difference
Measurement Sizes (length per side of square)Standard lens
Approx. 98 mm or more (depending on the distance)
Measurable Size for Typical Measurement Distance (size/distance)98mm / 250 mm Approx.

210mm / 500 mm Approx.

440mm / 1,000 mm Approx.

890mm / 2,000 mm Approx.

Measurable Luminance Range (including ND filter use)0.05 – 100,000 cd/m2
Measurement TimeSingle : Approx. 5 sec. or more; 4-time integration : Approx. 6 sec. or more; 16-time integration: Approx. 10 sec. or more; 64-time integration : Approx. 25 sec. or more; 256-time integration : Approx. 80 sec. or more
AccuracyLuminance±3%
Chromaticity±0.005
Temperature/Humidity Drift (within the operating temperature/ humidity range)
Luminance: ±2% of change compared to reference temperature and relative humidity of 23°C and 40%
Chromaticity: ±0.004 of change compared to reference temperature and relative humidity of 23°C and 40%
RepeatabilityLuminance0.5%
Chromaticity0.001
Inter-Point ErrorLuminance±2%
Chromaticity±0.002
Luminance±3%
Chromaticity±0.003
Other FunctionsMeasurement sync (Synchronization frequency selectable), User calibration Integration function
InterfaceUSB 2.0 or higher
Operating Temperature and Humidity Range10-30°C, Relative humidity 70% or less/No condensation
Storage Temperature and Humidity Range0-30°C, Relative humidity 70% or less/No condensation, 30-35°C, Relative humidity 55% or less/No condensation
SizeBody only160 (W) x 164 (H) X 192 (D) mm (Height including handle: 211 mm)
When lens and lens hood are attached223 (D) mm
Weight3.5 kg approx. (when standard lens and lens hood are attached)
Power SourceAC Adapter 100-240 V ~ , 1.2 A, 50-60 Hz

 

ModelCA-2500W
Light ReceptorCCD image sensor (monochrome) ; 2/3-inch ; Effective number of pixels : 1,000 x 1,000 pixels ; Equipped with XYZ filter (closely matches CIE 1931 color-matching function) and ND filter
LensInterchangeable Standard, wide, and telephoto lenses; low-magnification and high-magnification macro rings (for use with telephoto lens)
Measurement Points (Resolution)980 x 980 (490 x 490 or 196 x 196 selectable by using Data Management Software CA-S25w)
Color Indication ModesXYZ, LVxy, LVu’v’, T∆uv, Dominant wavelength, Excitation purity, LV contrast
Display ModesPseudocolor, Chromaticity diagram, Spot, Cross section, Color Difference
Measurement Sizes (length per side of square)Wide lens
Approx. 145 mm or more (depending on the distance)
Measurable Size for Typical Measurement Distance (size/distance)145mm / 200 mm Approx.

410mm / 500 mm Approx.

850mm / 1,000 mm Approx.

1,770mm / 2,000 mm Approx.

Measurable Luminance Range (including ND filter use)0.05 – 100,000 cd/m2
Measurement TimeSingle : Approx. 5 sec. or more; 4-time integration : Approx. 6 sec. or more; 16-time integration: Approx. 10 sec. or more; 64-time integration : Approx. 25 sec. or more; 256-time integration : Approx. 80 sec. or more
AccuracyLuminance±3%
Chromaticity±0.005
Temperature/Humidity Drift (within the operating temperature/ humidity range)
Luminance: ±2% of change compared to reference temperature and relative humidity of 23°C and 40%
Chromaticity: ±0.004 of change compared to reference temperature and relative humidity of 23°C and 40%
RepeatabilityLuminance0.5%
Chromaticity0.001
Inter-Point ErrorLuminance±2%
Chromaticity±0.002
Luminance±3%
Chromaticity±0.003
Other FunctionsMeasurement sync (Synchronization frequency selectable), User calibration Integration function
InterfaceUSB 2.0 or higher
Operating Temperature and Humidity Range10-30°C, Relative humidity 70% or less/No condensation
Storage Temperature and Humidity Range0-30°C, Relative humidity 70% or less/No condensation, 30-35°C, Relative humidity 55% or less/No condensation
SizeBody only160 (W) x 164 (H) X 192 (D) mm (Height including handle: 211 mm)
When lens and lens hood are attached219 (D) mm
Weight3.5 kg approx. (when standard lens and lens hood are attached)
Power SourceAC Adapter 100-240 V ~ , 1.2 A, 50-60 Hz

 

ModelCA-2500T
Light ReceptorCCD image sensor (monochrome) ; 2/3-inch ; Effective number of pixels : 1,000 x 1,000 pixels ; Equipped with XYZ filter (closely matches CIE 1931 color-matching function) and ND filter
LensInterchangeable Standard, wide, and telephoto lenses; low-magnification and high-magnification macro rings (for use with telephoto lens)
Measurement Points (Resolution)980 x 980 (490 x 490 or 196 x 196 selectable by using Data Management Software CA-S25w)
Color Indication ModesXYZ, LVxy, LVu’v’, T∆uv, Dominant wavelength, Excitation purity, LV contrast
Display ModesPseudocolor, Chromaticity diagram, Spot, Cross section, Color Difference
Measurement Sizes (length per side of square)Telephoto lens
Approx. 115 mm or more (depending on the distance)

*With low-magnification macro ring: Approx. 57mm (Fixed)

*With high-magnification macro ring: Approx. 27mm (Fixed)

Measurable Size for Typical Measurement Distance (size/distance)115mm / 900 mm Approx.

275mm / 2,000 mm Approx.

420mm / 3,000 mm Approx.

*With low-magnification macro ring: 57mm / 500 mm Approx. (Fixed)

*With high-magnification macro ring: 27mm / 300 mm Approx. (Fixed)

Measurable Luminance Range
(including ND filter use)
0.25 – 100,000 cd/m2

*With low-magnification macro ring: 0.25 – 100,000 cd/m2

*With high-magnification macro ring: 0.5 – 100,000 cd/m2

Measurement TimeSingle : Approx. 5 sec. or more; 4-time integration : Approx. 6 sec. or more; 16-time integration: Approx. 10 sec. or more; 64-time integration : Approx. 25 sec. or more; 256-time integration : Approx. 80 sec. or more
AccuracyLuminance±3%
Chromaticity±0.005
Temperature/Humidity Drift (within the operating temperature/ humidity range)
Luminance: ±2% of change compared to reference temperature and relative humidity of 23°C and 40%
Chromaticity: ±0.004 of change compared to reference temperature and relative humidity of 23°C and 40%
RepeatabilityLuminance0.5%
Chromaticity0.001
Inter-Point ErrorLuminance±2%
Chromaticity±0.002
Luminance±3%
Chromaticity±0.003
Other FunctionsMeasurement sync (Synchronization frequency selectable), User calibration Integration function
InterfaceUSB 2.0 or higher
Operating Temperature and Humidity Range10-30°C, Relative humidity 70% or less/No condensation
Storage Temperature and Humidity Range0-30°C, Relative humidity 70% or less/No condensation, 30-35°C, Relative humidity 55% or less/No condensation
SizeBody only160 (W) x 164 (H) X 192 (D) mm (Height including handle: 211 mm)
When lens and lens hood are attached224 (D) mm

*With low-magnification macro ring: 230 (D) mm

*With high-magnification macro ring: 237 (D) mm

Weight3.5 kg approx. (when standard lens and lens hood are attached)
Power SourceAC Adapter 100-240 V ~ , 1.2 A, 50-60 Hz

 

2D Color Analyzer CA-2500

Catalog Download: Download PDF (2.5MB)

Owner’s Manual: Download PDF (1.7MB)

 

2D Color Analyzer CA-2500



    All fields marked with an (*) are required.


    *Your data will be processed in accordance with our Privacy Policy.
    *Email us directly with any form submission problems