(Discontinued) CM-3700d Spectrophotometer
The CM-3700D Spectrophotometer is a bench top model that is precise and reliable. This device has a complete optical system that goes beyond industry standards along with aperture masks and a large sphere. Measurements are subjected to a high powered xenon flash to ensure the greatest level of accuracy. Depending on the application, users have the choice of switching between measuring areas; 25.4mm, 8mm, and 3x5mm.
The CM-3700D Spectrophotometer measures reflective colors with ease and can be switched from SCE and SCI effortlessly. This device’s UV cut filters allow the user to measure florescent colors via an automatic 1,000-step adjustment of UV light; this process provided excellent measurements of pulp, paper, and other florescent infused materials while transparent colors are effectively measured through diffused lighting. The impressive features of this device allow measurements to be taken form an array of materials including, but not limited to, textiles, plastics, kiln-fired materials, dyed items, food, and liquids.
The CM-3700D Spectrophotometer works well with Windows based computers via a USB connection. The device is compatible with a variety of Windows programs allowing for the easy editing and data management.
The CM-3700D Spectrophotometer is compatible the SpectrMagic NX software. The SpectraMagic software is user friendly and displays multiple, editable views of data results. Two or more secondary targets can be associated with targeted data from a primary target; this capability coincides with sophisticated QC applications like checking for color differences.
The software works on a drag-and drop platform which allows for easy interpretation after printing. For displaying information there is a variety of graph available along with preconfigured templates. There is also a status window in which users may view instrument information. Information compiled through older software is compatible with current software and can easily be transferred; all data is Excel exportable. Included with the software are a Color Tutorial and a Navigation Function to provide users with additional information.
(Discontinued) CM-3700d Spectrophotometer
Illumination/observation system | d/8 (diffused illumination; 8° viewing angle); SCI (specular component included)/SCE (specular component excluded) switchable; meets ISO and DIN standards for d/8 geometry; also meets CIE and ASTM standards for d/0 geometry. Transmittance: d/0 (diffuse illumination/0° viewing angle) |
Detector | Silicon photodiode array with flat holographic grating |
Wavelength range | 360nm to 740nm |
Wavelength pitch | 10nm |
Half bandwidth | Approximately 14nm average |
Photometric range | 0 to 200%; Resolution: 0.001% |
Light source | Pulsed xenon arc lamp |
Measurement time | 0.6 to 0.8 sec. (to start of data output) |
Illumination/measurement areas | Reflectance: Changeable between LAV, MAV, and SAV LAV: Φ28mm illumination/Φ25.4mm measurement MAV: Φ 11mm illumination/Φ8mm measurement SAV: 5 × 7mm illumination/3 × 5mm measurement Transmittance: Approx. Φ20mm |
Repeatability | When white calibration plate is measured 30 times at 10-sec. intervals after white calibration has been performed: ; Spectral reflectance: Standard deviation within 0.05% Chromaticity: Standard deviation within ΔE*ab 0.005 When black tile (BCRA Series II; reflectance: 1%) is measured 30 times at 10-sec. intervals after white calibration has been performed: Spectral reflectance: 380 to 740nm: Standard deviation within 0.02% 360 and 370nm: Standard deviation within 0.04% Chromaticity: Standard deviation within ΔE*ab 0.05 |
Inter instrument agreement | Mean ΔE*ab 0.08 (typical) Average for 12 BCRA Series II color tiles. Max ΔE*ab 0.3 (corresponds to ΔEcmc 0.2) for any of 12 BCRA Series II color tiles compared to values measured with master body |
Temperature drift | Spectral reflectance: Within ±0.10%/°C Color difference: Within ΔE*ab 0.05/°C |
UV adjustment | Computer controlled; continuously variable |
Specimen conditions for transmittance measurements | |
Interface | RS-232C standard; baud rates: 12,00, 2400, 4800, or 9600 |
Power | AC 100V/120V/230V 50/60Hz (using included AC adapter) |
Operating temperature/humidity range | 13 to 33°C, relative humidity 80% or less with no condensation |
Storage temperature/humidity range | 0 to 40°C, relative humidity 80% or less with no condensation |
Dimensions (W × H × D) | 271 × 259 × 500mm (10-11/16 × 10-3/16 × 19-11/16 in.) |
Weight | 18kg (39.7 lb.) |
Standard accessories | White Calibration Plate; Target Mask (3 × 5mm); Target Mask (Φ8mm); Target Mask (Φ25.4mm); Zero Calibration Box; AC Adapter; RS-232C Cable (2m/6.6 ft., 9-pin, for IBM PC); Accessory Case CM-66 |
Optional accessories | Color Data Software SpectraMagic NX; Transmittance Specimen Holder; Glass Cells (2mm, 10mm, 20mm); Plastic Cells (2mm, 10mm, 20mm); Transmittance Zero Calibration Plate; Color Tiles; RS-232C Cable (for IBM PC) |
Specifications subject to change without notice.