View our latest version of this instrument, the CA-410.

The Ca-210 color analyzer is a simple use device that measures the color and brightness oc LCD light displays quickly and with extreme accuracy.

(Discontinued) CA-210 Color Analyzer

The CA-210 Color Analyzer measures the color and brightness of LCD light displays accurately and easily. There are two types of probes available for this device each one offering cord lengths and measurement diameters. Users can select either the LCD flicker measuring probe or the universal measuring probe according to the task.

The CA-210 Color Analyzer is an impressively fast device capable of measuring 20 cycles per second. This device has a luminance measuring range starting at 0.1° cd/m2. The CA-210 uses XYZ filters for comparable measurements to the CIE 1931 Standard Observer. This device can measure distances as short as 20-40mm making it ideal for compact systems. Other advantageous features include the ability to connect up to 5 probes, a precise four digit color display, and a unique optical system with an acceptance angle of ±2.5°.

 

(Discontinued) CA-210 Color Analyzer

  • Quick measurements – 20 cycles per second.
  • Easy to use – simple design.
  • Highly accurate/Multi-function – extremely accurate calculations of even low luminances using either a universal or LCD flicker measuring probes.
  • Unique optical system and color display – four digit color display for greater precision.
  • Connect up to 5 sensors – using the expansion board attachment, sold separately, have the capability to utilize up to 5 probes.

(Discontinued) CA-210 Color Analyzer

 

ReceptorSilicon photocell
Measurement areaΦ27mm
Acceptance angle±2.5°; Pointing function: By LED
Measurement distance30±10mm
Display rangeLuminance0.01 to 1000cd/m2
ChromaticityDisplayed in 4- or 3-digit value (selectable)
LuminanceMeasurement range0.10 to 1000cd/m2
Accuracy (for white)*1±2%, ±1 digit of reading (Temperature: 23°C ±2°C; Relative humidity:(40±10)%)
Repeatability*10.10 to 0.99cd/m2: 0.2%+1digit; 1.0 to 1000cd/m2: 0.1%+1digi
ChromaticityMeasurement range0.10 to 1000cd/m2
Accuracy*10.10 to 4.99cd/m2: ±0.008 for white5.00 to 39.99cd/m2: ±0.005 for white40.00 to 1000 cd/m2: ±0.003 for white160 cd/m2: ±0.002 for white; ± 0.004 for monochrome *2
Repeatability*10.10 to 0.19cd/m2: 0.0150.20 to 0.49cd/m2: 0.0080.50 to 1.99cd/m2: 0.0032.00 to 1000 cd/m2: 0.001
Measurement speed*3xyLv0.10 to 1.99 cd/m2: 5 (4.5) 4 cd/m2 or greater: 20 (17)
DisplayDigitalxyLv, XYZ, TΔuvLv, u’v’Lv, RGB analyze
ΔxΔyΔLv, R/G B/G ΔG, ΔR B/R G/R
LCD16 characters × 2 lines (with backlight)
SYNC modeNTSC, PAL, EXT, UNIV, INT
Measurement subjectVertical synchronous frequency: 40 to 200Hz
Memory channels100
Analyzer functionStandard function
InterfaceUSB (Rev. 1.1) or RS-232c (38,400 bps maximum)
Multi-point measurement Max.5 points (using optional 4-Probe Expansion Board CA-B14)
Software SolutionsSDK software (supplied as standard accessory)
Operating temperature/humidity rangeTemperature: 10 to 28°C; Relative humidity 70% or less with no condensation. Luminance change: ±2% ±1 digit of reading for white; Chromaticity change: ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta’s standard LCD *1, 160.0 cd/m2 at 23°C 40%RH
Storage temperature/humidity range0 to 28°C: Relative humidity 70% or less with no condensation; 28 to 40°C: Relative humidity 40% or less with no condensation
Input voltage range100 to 240V~50/60Hz, 50VA
Dimensions (W × H × D)Main body: 340 × 127 × 216mmProbe: Φ49 × 204mm
WeightMain body: 3.58kg; Probe: 520g

 

(Discontinued) CA-210 Color Analyzer