The LS-150 (successor to LS-100) possess a minimum measuring area of 14.4mm and a greatly improved luminance range of up to 999,900 cd/m². It is a highly accurate and easy-to-use luminance meter that conform to DIN 5032-7 Class B.

Luminance Meter LS-150

The LS-150 Luminance Meter (successor to the LS-100), with compact and ergonomic design, is a highly accurate luminance meter with a spectral response that closely matches the V(λ) spectral luminous efficiency function of the human eye, providing measurement results that correlate with visual evaluation.

The LS-150 uses a single lens reflex (SLR) optics system which provide precise control over targeting and ensures that the viewfinder shows the exact area to be measured, even at very short distances. Areas as small as 1.3mm in diameter (when using optional close-up lenses) can be also be measured accurately. The value of the measured luminance is displayed both inside the viewfinder and the external monochrome display.

The LS-150 conform to DIN 5032-7 Class B ans is capable of measuring luminance as low as 0.001 cd/m² and as high as 999,900 cd/m². LS-150 also comes with color-correction and calibration functions bthat allow users to choose a standard luminance and have differences corrected according to any light sources specified.

pictureTwo-fix-2

Luminance Meter LS-150

  • 1° measurement angle
  • conform to DIN 5032-7 Class B
  • Luminance measurement range of 0.01 to 999,900 cd/m2
  • Portable, compact and battery powered(2xAA)
  • SLR (Single Lens Reflex), precise targeting, showing exact area to measure
  • Easy-to-read backlit display for viewing in dark areas
  • Includes Utility Software which allows instrument to be controlled from the PC via USB 2.0
  • Optional CCD camera adapter which enables viewfinder to be monitored from distance

Luminance Meter LS-150

Model LS-150 Luminance Meter
Measuring angle
Optical system SLR viewing system, f = 85 mm F2.8
Angle of view 9° (with diopter adjustment)
Relative spectral responsivity Closely matches spectral luminous efficiency  (V (Λ))
Application standard DIN 5032-7 Class B compliant
Minimum measuring area (diameter) 14.4 mm
(1.3 mm when close-up lens is used)
Minimum measuring distance (From the measuring distance reference plane) 1,012 mm
(213 mm when close-up lens is used)
Measurement mode Instantaneous value, maximum/minimum value, luminance difference (Δ)/luminance ratio (%)
Measurement time Auto: 0.7 to 4.3 seconds Manual: 0.7 to 7.1 seconds
Luminance unit cd/m2 or fL
Luminance range 0.001 to 999,900 cd/m2
Accuracy*1 (1 cd/m2 or less) +/-2% +/- 2 digit
(1 cd/m2 or more) +/-2% +/- 1 digit
Repeatability*1 0.2% + 1 digit
Calibration standard Konica Minolta standard/user-specified standard switchable
User calibration channels 10 channels
Data memory 1,000 data
External display
(Number of significant digits)
(Luminance) 4 digits
(Chromaticity) 4 digits
Internal display
(Number of significant digits)
(Luminance) 4 digits
Interface USB 2.0
Power AA-size batteries (x2), USB bus power, or optional AC adapter
Current consumption When viewfinder display is lit: 70 mA average
Operation temperature/humidity range 0 to 40°C, relative humidity of 85% or less (at 35°C)
Storage temperature/humidity range 0 to 45°C, relative humidity of 85% or less (at 35°C)
Size 71x214x154 mm
Weight 850 g (without batteries)
Standard accessories Lens Cap
Eyepiece ND Filter
Eyepiece Cap
AA-size batteries (x2)
Hard Case CS-A12
Wrist Strap CS-A13
USB Cable T-A15
Data Management Software CS-S20
Optimal accessories Close-Up Lens No. 153/135/122/110
CCD Camera Adapter CS-A14
Illuminance Adapter CS-A15
AC Adapter AC-A305J/L/M
Specifications are subject to change without notice.
*1 Standard Illuminant A; Standard measurement distance; Measurement time setting: Auto

Luminance Meter LS-150

Catalog Download: Download PDF (3.5MB)

Owner’s Manual: Download PDF (6.2MB)

 

Luminance Meter LS-150



    All fields marked with an (*) are required.


    Email us directly with any form submission problems